Pulse height analysis and pulse shape discrimination of pure LaCl3 scintillation crystal across a broad neutron energy range

บทความในวารสาร


ผู้เขียน/บรรณาธิการ


กลุ่มสาขาการวิจัยเชิงกลยุทธ์


รายละเอียดสำหรับงานพิมพ์

รายชื่อผู้แต่งSangaroon, S.; Kim, H.J.; Quang, N.D.; Ogawa, K.; Isobe, M.; Kaewkhao, J.; Wantana, N.; Wisitsorasak, A.; Toyama, S.; Miwa, M.; Matsuyama, S.; Intachai, N.; Kothan, S.

ผู้เผยแพร่Elsevier

ปีที่เผยแพร่ (ค.ศ.)2026

วารสารRadiation Physics and Chemistry (0969-806X)

Volume number239

หน้าแรก113290

นอก0969-806X

eISSN1879-0895

URLhttps://www.scopus.com/inward/record.uri?eid=2-s2.0-105016631985&doi=10.1016%2Fj.radphyschem.2025.113290&partnerID=40&md5=c3580acf86dfa732d0e7e69ff4f3df4f

ภาษาEnglish-Great Britain (EN-GB)


ดูบนเว็บไซต์ของสำนักพิมพ์


บทคัดย่อ

A pure LaCl3 crystal scintillation detector has been developed and shows strong potential for fast neutron detection, particularly via the 35Cl(n,p)35S and 35Cl(n,α)32P reactions. These reaction channels make it a promising candidate for neutron spectroscopy applications in nuclear fusion research. In this study, a comprehensive characterization of the pure LaCl3 scintillation detector was performed in a mixed neutron and γ-ray radiation field. Pulse height analysis and neutron–γ pulse shape discrimination were systematically evaluated over a wide neutron energy range, from approximately 2.46 MeV to 16.89 MeV, using both a 252Cf spontaneous fission source and a mono-energetic neutron beam at the Fast Neutron Laboratory, Tohoku University. The results reveal a clear separation between neutron- and γ-induced signals, with figure-of-merit values exceeding 1.18 at specific energies, demonstrating the detector's excellent pulse shape discrimination performance. The pulse height response associated with charge generation in the pure LaCl3 scintillation crystal was systematically characterized. Energy calibration and linearity for neutron energies up to 5.61 MeV were verified through the peak corresponding to the 35Cl(n,p)35S ground-state reaction. These findings confirm that pure LaCl3 scintillators exhibit excellent neutron–γ discrimination, enabled by differences in the pulse height and decay time characteristics of the induced signals, making them well-suited for high-resolution neutron spectroscopy in mixed radiation fields typical of magnetic confinement fusion environments. © 2025 Elsevier Ltd


คำสำคัญ

ไม่พบข้อมูลที่เกี่ยวข้อง


อัพเดทล่าสุด 2026-17-02 ถึง 12:00