Heart sound monitor for biomedical instrumentation

Conference proceedings article


Authors/Editors


Strategic Research Themes

No matching items found.


Publication Details

Author listSutdipong S., Tunlasakun K.

PublisherTrans Tech Publications

Publication year2013

Volume number303-306

Start page650

End page653

Number of pages4

ISBN9783037856529

ISSN1660-9336

eISSN1662-7482

URLhttps://www.scopus.com/inward/record.uri?eid=2-s2.0-84874730140&doi=10.4028%2fwww.scientific.net%2fAMM.303-306.650&partnerID=40&md5=d6c0afdd4bbb3bd2cffbec62f208abfd

LanguagesEnglish-Great Britain (EN-GB)


View in Web of Science | View on publisher site | View citing articles in Web of Science


Abstract

This research presents the design and development of the heart sound monitor for biomedical instrumentation which can be worked with a personal computer. The prototype will receive the heart sound via the condenser microphone built-in the stethoscope. The condenser microphone will be conversed the air pressure from heart beats to electrical signal that signal will transformed to computer via sound card. The sound card will be conversed the analog signal to digital signal for process by heart sound processing program developed by LabVIEW program. The signal will be analyzed with short time Fourier transforms in heart sound processing program by graphical user interface. The user is able to select a band pass of signal for filter and choose the frequency spectrum of heart sound for display. The output database from this prototype is necessary for Medical Education or Clinical Practice. ฉ (2013) Trans Tech Publications, Switzerland.


Keywords

Heart beatsHeart soundStethoscope


Last updated on 2023-29-09 at 07:35