Lead Assays with Smartphone Detection Using a Monolithic Rod with 4-(2-Pyridylazo) Resorcinol

Journal article


Authors/Editors


Strategic Research Themes


Publication Details

Author listPiyanat Issarangkura Na Ayutthaya, Chonnipa Yeerum, Kullapon Kesonkan, Kanokwan Kiwfo, Kate Grudpan, Norio Teshima, Hiroya Murakami and Monnapat Vongboot

PublisherMDPI

Publication year2021

Volume number26

Issue number18

Start page5720

End page-

ISSN1420-3049

eISSN1420-3049

URLhttps://www.mdpi.com/1420-3049/26/18/5720

LanguagesEnglish-United States (EN-US)


View on publisher site


Abstract

The monolithic rod of polyurethane foam-[4-(2-pyridylazo) resorcinol] (PUF-PAR) as a simple chemical sensor for lead assay with a smartphone detection and image processing was developed. With easily available simple apparatus such as a plastic cup and a stirrer rod, the monolithic PUF rod was synthesized in a glass tube. The monolithic PUF-PAR rod could be directly loaded by standard/sample solution without the sample preparation. A one-shot image in G/B value from profile plot in ImageJ for a sample with triplicate results via a single standard calibration approach was obtained. A linear single standard calibration was: [G/B value] = −0.038[µg Pb2+] + 2.827, R2 = 0.95 for 10-30 µg Pb2+ with a limit of quantitation (LOQ) of 33 µg L−1. The precision was lower than 15% RSD. The proposed method was tested by an assay for Pb2+ contents in drinking water samples from Bangkok. The results obtained by the proposed method agree with those of ICP-OES and with 100-120% recovery, demonstrating that the method is useful for screening on-site water monitoring.