Lead Assays with Smartphone Detection Using a Monolithic Rod with 4-(2-Pyridylazo) Resorcinol
Journal article
Authors/Editors
Strategic Research Themes
Publication Details
Author list: Piyanat Issarangkura Na Ayutthaya, Chonnipa Yeerum, Kullapon Kesonkan, Kanokwan Kiwfo, Kate Grudpan, Norio Teshima, Hiroya Murakami and Monnapat Vongboot
Publisher: MDPI
Publication year: 2021
Volume number: 26
Issue number: 18
Start page: 5720
End page: -
ISSN: 1420-3049
eISSN: 1420-3049
URL: https://www.mdpi.com/1420-3049/26/18/5720
Languages: English-United States (EN-US)
Abstract
The monolithic rod of polyurethane foam-[4-(2-pyridylazo) resorcinol] (PUF-PAR) as a simple chemical sensor for lead assay with a smartphone detection and image processing was developed. With easily available simple apparatus such as a plastic cup and a stirrer rod, the monolithic PUF rod was synthesized in a glass tube. The monolithic PUF-PAR rod could be directly loaded by standard/sample solution without the sample preparation. A one-shot image in G/B value from profile plot in ImageJ for a sample with triplicate results via a single standard calibration approach was obtained. A linear single standard calibration was: [G/B value] = −0.038[µg Pb2+] + 2.827, R2 = 0.95 for 10-30 µg Pb2+ with a limit of quantitation (LOQ) of 33 µg L−1. The precision was lower than 15% RSD. The proposed method was tested by an assay for Pb2+ contents in drinking water samples from Bangkok. The results obtained by the proposed method agree with those of ICP-OES and with 100-120% recovery, demonstrating that the method is useful for screening on-site water monitoring.