Determination of Lead Employing Simple Flow Injection AAS with Monolithic Alginate-Polyurethane Composite Packed
In-Valve Column
Journal article
Authors/Editors
Strategic Research Themes
Publication Details
Author list: Piyanat Issarangkura Na Ayutthaya, Chonnipa Yeerum, Kullapon Kesonkan, Kanokwan Kiwfo,
Kate Grudpan,*, Norio Teshima, Hiroya Murakami and Monnapat Vongboot*
Publisher: MDPI
Publication year: 2021
Volume number: 26
Issue number: 15
Start page: 4397
ISSN: 1420-3049
eISSN: 1420-3049
URL: https://www.mdpi.com/1420-3049/26/15/4397
Abstract
A simple flow injection FlameAAS for lead determination with an alginate-polyurethane
composite (ALG-PUC) monolithic in-valve column has been developed. The ALG-PUC monolithic
rod was prepared by mixing methylene diphenyl diisocyanate with polyol and sodium
alginate with the ratio of 2:1:1 by weight for a 5 min polymerization reaction. It was then put
into a column (0.8 cm i.d x 11 cm length) situated in a switching valve for the FI set up. A single
standard calibration could be obtained by plotting the loaded ug Pb2+ vs. FI response (absorbances).
The loaded ug Pb2+ is calculated: ug Pb2+ = FRload x LT x CPb2+, where the FR load is the flow rate of the loading analyte solution (mL min1), LT is the loading time (min), and CPb2+ is the Pb2+ concentration (ug mL-1). A linear calibration equation was obtained: FI response (absorbances) = 0.0018 [g Pb2+] + 0.0032, R2 = 0.9927 for 1–150 ug Pb2+, and RSD of less than 20% was also obtained. Application of the developed procedure has been demonstrated in real samples.