Real-Time Double-Layer Thin Film Thickness Measurements Using Modified Sagnac Interferometer with Polarization Phase Shifting Approach

Journal article


Authors/Editors


Strategic Research Themes


Publication Details

Author listAbdullahi Usman, Apichai Bhatranand, Yuttapong Jiraraksopakun, Rapeepan Kaewon and Chutchai Pawong

PublisherMDPI

Publication year2021

Volume number8

Issue number12

Start page1

End page13

Number of pages13

ISSNISSN 2304-6732

eISSN2304-6732

URLhttps://www.mdpi.com/2304-6732/8/12/529

LanguagesEnglish-United States (EN-US)


View on publisher site


Abstract

This paper describes a modified Sagnac interferometer with a self-referenced polarization
and phase-shifting technique for real-time thickness measurement of single- and double-layer
transparent thin films. The proposed interferometric setup generated outstanding rotating linearly
polarized light with a degree of polarization (DOP) of 99.40%. A beam splitter placed at the interferometer
output separated the beam into two identical linearly polarized beams. One of the beams
served as a reference, while the other served as a sensing arm. The output linear polarizer set at
45 relative to a reference plane was positioned anterior to the photodetectors to get rotating light
intensities for phase shift measurement; hence, the intensities at various polarizations of 0, 45,
and 90 were automatically acquired without any polarizing device adjustments. These intensities
were then transformed into a phase retardation introduced by a sample, and the resulting phase
shift was eventually converted into film thickness. The samples were properly prepared, with pure
BK7 substrate being deposited by WO3-, Ta2O5-, and WO3/Ta2O5 films of known thicknesses. The
thickness measurement obtained from the proposed system yielded reading errors of 1.3%, 0.2%, and
1.3/2.5% for WO3-, Ta2O5-, and WO3/Ta2O5 films, respectively. The mathematical theory was effectively
demonstrated and empirically confirmed. The experimental results show that the proposed
setup has a lot of potential for real-time, non-destructive thickness assessment of transparent thin
films without the need to modify polarizing device orientations.